A laminar flow pass-through box is an auxiliary cleanroom device used to transfer small items between a cleanroom and a non-clean area, or between cleanrooms of different cleanliness classes or pressure levels. It functions as an airlock, preventing pressure loss in higher-grade cleanrooms, while simultaneously providing a self-cleaning effect during the transfer process. This ensures that items entering the clean zone are clean, thereby minimizing cross-contamination. The airflow (purging) duration can be adjusted via a manual switch, optimizing both self-cleaning performance and energy efficiency.
I. Installation of the Laminar Flow Pass-Through Box
Install and operate in strict accordance with established standards.
The unit requires a 220V/50Hz power supply. The installation environment should have a temperature of 15°C–35°C and a relative humidity (RH) of 40%–70%.
Disconnect the power supply during installation to prevent electric shock.
Ensure safety during wiring and verify that the unit’s chassis is properly grounded.
Take care during installation to avoid contact with other objects, which could scratch the unit’s housing.
Adhere strictly to the specified environmental requirements; failure to do so may compromise the unit’s performance, lead to short circuits, damage electrical components, and shorten the unit's service life.
Install the unit away from heaters, stoves, or other heat sources.
Do not place flammable materials in the installation area, as this poses a fire risk.
The use of explosive or volatile chemicals is strictly prohibited, as the unit may draw these substances in, potentially causing fires or explosions.
Do not install the unit in environments containing sulfuric acid, hydrochloric acid, or other corrosive chemicals, as this could lead to corrosion and damage.
After installation, wipe the interior and exterior of the unit clean.
Recommended filter replacement cycle: Replace the filter when its operating resistance reaches twice its initial resistance.
II. Pre-Operation Checks
Verification of Filtration Efficiency and Airflow Velocity
Ensure the external power supply is correctly connected and functioning properly.
Check that the return air grille is not obstructed; blockages can prevent the unit from operating correctly, potentially causing malfunctions or damage to electrical components. Operate the laminar flow pass box according to the provided instructions. The fan should function as scheduled, operating without any noticeable abnormal noise or vibration.
The unit features an interlock system for the two doors (i.e., when one door is open, the other cannot be opened). The germicidal lamp should turn on and off according to the set timer.
Integrity testing for the laminar flow pass box filter: Use a PAO aerosol photometer and an aerosol generator to test the liquid-seal HEPA filter. The test should be conducted at a distance of 25.4 mm from the air outlet surface, with a scanning speed not exceeding 5 mm/s. The leakage rate must be less than 0.01% (the unit must be powered on and allowed to self-clean for at least 10 minutes prior to testing).
Air velocity testing for the laminar flow pass box: Use an anemometer to measure the air velocity at a distance of 100–150 mm below the air outlet surface; the average air velocity should be between 0.35 and 0.65 m/s.
III. Precautions for Use
Operate the unit strictly in accordance with the established specifications.
Wiring must be performed according to the provided wiring guide to prevent incorrect port connections. Always disconnect the power supply before performing any wiring or maintenance.
Do not use chemical solvents (such as paint thinners) to clean the stainless steel surfaces.
The power board utilizes digital circuitry; keep it away from high-frequency or high-current devices to prevent system freezes.
Ensure no foreign objects enter the ports, as this could cause a short circuit.
Do not operate the control buttons on both sides simultaneously, as this may cause the system to freeze. If a freeze occurs, simply power the unit off and back on.
Do not use compressed air to clean the HEPA filter at close range.
Since the ultraviolet (UV) light from the germicidal lamp can penetrate and destroy cells, ensure that the UV light does not shine directly onto human skin or eyes. Do not look directly at the lamp while it is illuminated. Because short-wave UV light cannot pass through ordinary glass, wearing eyeglasses can help protect your eyes. If accidental eye exposure occurs, it is generally not a major concern—similar to a sunburn—but in more severe cases, eye drops or human breast milk may be used to aid recovery. When replacing the germicidal lamp, please disconnect the power supply and allow the lamp to cool down before proceeding, in order to prevent burns from high temperatures.
IV. Maintenance and Care
Maintenance Requires Professional Training
Daily maintenance and care of the laminar flow pass box must be performed by personnel who have received professional training. If any irregularities are detected, disconnect the power supply immediately and take corrective action.
Daily Maintenance and Inspection
01: Fan and Motor
The centrifugal fan must rotate smoothly without producing abnormal noise or vibration. If abnormal noise is heard, check for foreign objects inside the fan unit and inspect the motor bearings for unusual sounds. If the motor is faulty, replace it.
02: Germicidal Lamp
The germicidal lamp should function properly, with no flickering or unexpected shut-off.
03: Control Panel
The buttons on the control panel should function correctly, and all control adjustments should operate as intended.
04: Differential Pressure Gauge
The differential pressure gauge should display normal readings, and the differential pressure must not exceed the specified final resistance value (a final resistance of 400 Pa – 500 Pa is recommended).
05: Exterior of the Unit
Check the exterior of the unit for any signs of scratches or corrosion.
Q: Periodic Motor Inspection
A: Criteria:
Check the motor for unusual noise, vibration, or odors; inspect the bearings for abnormal sounds.
Check cables and terminals for damage or corrosion.
Signs and Potential Causes:
Bearing damage may cause the motor to run unevenly or seize.
Poor insulation at power connections or terminals may lead to abnormal heating, grounding faults, or short circuits.
Q: Periodic Fan Inspection
A: Criteria:
Inspect the centrifugal fan for abnormal noise, vibration, or damage; replace any faulty components.
Signs and Potential Causes:
Accumulated dust or physical damage can cause operational imbalance, leading to further damage.
Q: Periodic Inspection of the Outer Casing and Other Components
A: Criteria:
Visually inspect components for defects, such as discoloration or scratches. Signs and potential causes:
Corrosion or rusting of components due to chemical agents, resulting in the generation of dust.
Q: Periodic inspection of other electrical components and wiring materials.
A: Acceptance criteria:
Check for signs of corrosion, exposed conductors, or damage to wire insulation.
Signs and potential causes:
Poor insulation, loose terminal screws, or other issues leading to overheating, short circuits, or burning.
Q: Periodic PAO (Polyalphaolefin) testing.
A: Acceptance criteria:
1. Prior to testing, the laminar flow pass box must be fully operational and functioning correctly in all aspects, meeting its specified technical performance requirements.
2. Ensure that the testing equipment and all connections between the equipment and the unit are free from smoke leaks; prevent smoke from entering the working area of the laminar flow pass box, as this could compromise test results (if leaks are unavoidable, appropriate protective measures must be taken).
3. Once the laminar flow pass box is operating stably (after approximately 3 minutes), activate the PAO aerosol generator. Use a calibrated photometer to measure the PAO concentration inside the unit; once the concentration reaches ≥10 µg/L, proceed with the test according to the instrument’s operating instructions.
4. During testing, hold the sampling probe approximately 25 mm from the air outlet surface and move it at a speed not exceeding 5 cm/s to check for leaks in the HEPA filter and its mounting frame.
5. The leakage rate must be less than 0.01%.
Signs and potential causes:
Leakage resulting from damage to the filter.
Q: Static Airborne Particle Count Test
A: Acceptance Criteria:
1. Under static conditions, use calibrated instrumentation to measure airborne particles with diameters ≥0.5 µm and ≥5.0 µm within the laminar flow pass box.
2. Prior to testing, ensure the laminar flow pass box is fully operational and functioning correctly in all aspects, in compliance with its technical performance specifications.
3. Turn on the laminar flow pass box, allow the fan to run for approximately 3 minutes, and position the testing instrument at the designated sampling points.
4. Sampling points shall be located in the area beneath the filter; a total of 2 sampling points are required.
Sampling frequency: One sample per point.
Sample volume: A minimum of 1 m³ per sampling point; record the actual sample volume and the corresponding instrument reading.
Indications and Potential Causes:
Leakage resulting from a damaged filter.
Q: UV Lamp Irradiance Test
A: Acceptance Criteria:
1. After the UV lamp has been on for 5 minutes, position the probe of a UV irradiance meter (calibrated for a wavelength of 254 nm) directly beneath the lamp at a vertical distance of 1 meter. Once the reading on the meter stabilizes, the displayed value represents the irradiance of the UV lamp.
2. Note the following conditions during testing: supply voltage of 220 V, temperature between 20°C and 25°C, and relative humidity below 60%. The UV irradiance meter must be within its valid calibration period as certified by the relevant metrological authority.
A value of ≥70 µW/cm² is considered acceptable.
Indications and Potential Causes:
The UV lamp has exceeded its service life, resulting in insufficient germicidal irradiance and compromised sterilization effectiveness.
Q: Noise Test
A: Criteria:
Measure the equipment's noise level using a calibrated sound level meter at a distance of 1.2 meters from the unit.
With the equipment turned off, measure the ambient noise level (background noise).
If the difference between the measured value and the background noise is 9 dB(A) or less, apply a correction; the corrected value shall be taken as the final measured result.
Apply corrections according to the following guidelines:
A) If the difference between the measured value and the background noise is less than 3 dB(A), the test is invalid.
B) If the difference between the measured value and the background noise is 3 dB(A), the correction value is -3.
C) If the difference between the measured value and the background noise is 4–6 dB(A), the correction value is -2.
D) If the difference between the measured value and the background noise is 7–9 dB(A), the correction value is -1.
E) If the difference between the measured value and the background noise is 10 dB(A) or greater, no correction is applied.
Standard: Noise level must be less than 65 dB.
Symptoms and Potential Causes:
Excessive noise caused by a fan malfunction.
V. Handling of Anomalies
Electrical Function, Cleanliness, and Airflow Velocity Testing
01: Electrical Function Anomalies
In the event of an equipment malfunction, you may attempt repairs yourself by following the instructions in the user manual, or contact our customer service department to request an on-site repair. For products within the warranty period, our company will not charge any repair fees (excluding damage caused by human error or misuse). For products beyond the warranty period, our company will charge fees based on the cost of any replacement parts used. (Note: Unauthorized modifications to the electrical circuitry are not covered under warranty. Filters and lighting fixtures are considered consumables and are also excluded from the warranty.)
02: Abnormal Cleanliness Test Results
If cleanliness levels fail to meet technical specifications, inspect the filter media for damage and ensure the filter clamping blocks are securely tightened. If these checks are satisfactory but leakage persists, wipe the filter frame with alcohol and re-verify. If the filtration efficiency still fails to meet requirements, the HEPA (High-Efficiency Particulate Air) filter must be replaced.
03: Abnormal Airflow Velocity Test Results
If the average airflow velocity of the Fan Filter Unit (FFU) is measured below 0.36 m/s, first verify that the fan is operating correctly. If the fan is malfunctioning, you may attempt repairs according to the instruction manual or contact our customer service department to request an on-site repair. If the fan is operating normally, the primary filter or the HEPA filter should be replaced.
Procedure for Replacing HEPA Filters:
a. Disconnect the power supply and prepare the replacement HEPA filter.
b. Use a wrench to remove the screws securing the diffuser panel on the air-outlet side of the laminar flow pass box, then remove the panel.
c. Use a wrench to remove the screws securing the HEPA filter clamping blocks inside the laminar flow pass box, then remove the filter.
d. Before installing the new filter, unpack it and perform a visual inspection—check for damage to the filter media, as well as any deformation or gaps in the frame. Secure the filter using the clamping blocks and nuts; do not overtighten.
e. After replacing the HEPA filter, verify that there is no leakage around the frame and conduct a PAO (polyalphaolefin) integrity test to ensure compliance with technical specifications. The unit is ready for normal operation once the average airflow velocity is maintained within the range of 0.45 m/s ± 20%.